High frequency electrosurgical unit that can be used to simultaneously perform monopolar and bipolar surgery.
DOUBLE OPERATIONAL ACTIVITY
- two output channels for monopolar cutting
- two independent calibrations and channels for the most frequent monopolar coagulations
CONTROLLED CUTTING PROCEDURES
- intensified cutting currents for incisions without eschar
- cutting current with coagulating properties for dissection with controlled bleeding
- low capacitive coupling in endoscopic procedures using Slow BLEND current
DOUBLE CONTROLLED COAGULATION PROCEDURES
- SPEEDY coagulation and DEEP coagulation available to 2 operators
- SPRAY coagulation obtainable with handpiece or pedal
BROAD BIPOLAR FEATURES
- bipolar cutting with selectable bleeding control
- slow BLEND cutting currents
- automatic start and stop of coagulation that can be activated separately
TOTAL SAFETY FOR PATIENT AND SURGEON
- acceptability of the contact impedance between the patient’s body and the neutral plate continuously verified by the Skin Plate Electronic Control (SPEC)
- continuous monitoring of radio frequency leakage current
- variations in mains voltage automatically compensated before starting the supply of power and radio frequency
CUSTOMIZED PROCEDURES
- independent choice of coagulation currents and power levels of each of the two operators
- ten personal procedures that can be configured and immediately recalled
CONDUCTION OF EXTENSIONS AND CONNECTION SIGNALS
- outputs for piloted connections of argon plasma units and fume extractors
AUTOMATIC SELF-DIAGNOSIS
- automatic diagnostic check of the complete current functionality of the electronic unit and connected accessories and coded indication of the malfunctions found.
- OUTPATIENT SURGERY
- GENERAL SURGERY
- PEDIATRIC SURGERY
- PLASTIC SURGERY
- THORACIC SURGERY
- VASCULAR SURGERY
- DERMATOLOGY
- ENDOSCOPY
- GASTROENTEROLOGY
- GYNECOLOGY
- NEUROSURGERY
- ORTHOPEDICS
- ENT
- PNEUMOLOGY
- UROLOGY
- VETERINARY
Technical features
• Max pure monopolar cutting power CUT: 300 W – 300 Ohm
• Max monopolar cutting-coag power CUT80%: 250 W – 300 Ohm
• Max monopolar cutting-coag power CUT60%: 200 W – 300 Ohm
• Max monopolar cutting-coag power CUT40%: 150 W – 300 Ohm
• Max ENHANCED monopolar cutting power: 200 W – 500 Ohm
• Max monopolar BLEND cutting-coag power: 200 W – 300 Ohm
• SPEEDY monopolar coagulation max power: 100 W – 500 Ohm
• Max DEEP monopolar coagulation power: 100 W – 200 Ohm
• Max SPRAY monopolar coagulation power: 70 W – 2000 Ohm
• Max monopolar coagulation power SPRAY ARGON: 70 W – 2000 Ohm
• Max bipolar CUT cutting power: 95 W – 150 Ohm
• Max bipolar cutting-coag cutting power CUT80%: 95 W – 150 Ohm
• Max bipolar cutting-coag cutting power CUT60%: 95 W – 150 Ohm
• Max bipolar cutting-coag cutting power CUT40%: 60 W – 150 Ohm
• Max bipolar cutting-coag power BLEND: 90 W – 150 Ohm
• COAG bipolar coagulation max power: 70 W – 100 Ohm
• Monopolar/bipolar working frequency: 425 kHz / 525 kHz
• Neutral electrode: F
• Selectable power supply: 115-230 Vac
• Mains frequency: 50-60 Hz
• Maximum absorbed power: 1000 VA
• Dimensions WxHxD: 470 x 150 x 400 mm
• Weight 17.5 kg
• Controls: OC plate/patient circuit control, Control of emitted power, Self-diagnosis
• Safety: EN60601-1, EN60601-1-2, EN60601-2-2
• Electrical Class: I CF, MDD 93/42/EC Class: II b
Standard equipment
1 – 5 m power cable
1 – Autoclavable handpiece with two buttons (30536)
5 – Sterile handpiece cold with two buttons (30551)
1 – Plate cable (30559)
5 – Rem disposable plates (30493)
2 – HP double watertight pedal set
3 – Short blade electrode 7 cm
3 – Blade electrode 16 cm long
3 – Short ball electrode Ø 4 mm, 6 cm
1 – Electrode cleaning sponge
1 – User manual (IT, GB)
Reviews
There are no reviews yet.