Description :
Neonatal incubator is warming equipment used for providing constant temperature and humidity for treatment of premature infants and critical neonates on the basis of satisfying their physiological characteristics and needs.
COMEN B3 incubator was thoughtfully designed to provides stable temperature and suitable humidity ambient, in addition to configure vital signs monitoring function to promoting the growth and stability of even the most fragile of patients.
Standard Configuration: Humidity Control, Temp (body temperature& incubator temperature), O2 Monitoring,
Optional Configuration: Weight Scale, Apnea Monitoring, Apnea awaking, Servo oxygen supply, Damping Doors
Safety Standards: ISO 13485:2003 approved, MDD 93/42/EEC as amended by MDD 2007/47/EC
Operation Environment
- Working Temp 20~30°C
- Air flow: ≤0.3m/s
- Humidity ≤80%
- Power Supply 100-240V~, 50/60Hz±1Hz
- Battery Type Rechargeable Lithium-ion battery
- Battery Capacity 4400mAh
- Battery Recharging Time: Maximum 6 hours for charging; Battery backup 1 hours for continuous working (such functions as heating, humidifying shall be excluded)
- Indicator: One alarm indicator, Power indicator, Battery indicator, QRS beep and alarm sound, Operating key sound
- Interface: Parameter cable interface, AC power input, Three auxiliary power output, Plug-in slot, USB port
- Date storage:
- Alarm Review: 200 groups
- Power-off storage: Yes
- Alarm: User-adjustable High and Low 3-level Limits; Prioritized audible and visual alarm
- Network: Connected to Central Monitoring System by hardwire/wireless
- Temperature Control
- Setting range: 35.0~37.5°C (Infant mode); 25.0~39.0°C (Air Mode), Set temperature exceeds 37°C, should press the display the “> 37°C” key, and then confirmed by the interface pop-up window to enter
- Measurement range: 25.0~45.0°C
- Resolution: 0.1°C
- Accuracy: ±0.3°C
- Alarm range: 34~38.5°C (Infant mode), 23~41°C (Air Mode)
- Humidification time: ≤50mins (In indoor environment to 25°C±2°C, relative humidity of 40%±10% conditions, reach to 36°C, relative humidity of 85%)
- Heating time: ≤35mins (Temperature inside the thermostat cover rises by 11°C)
- Operation without adding water time: ≥12 hours
- Humidity Control
- Relative humidity range: 30%~95%
- Resolution: 1%
- Accuracy: ±10% (Humidity less than 85%); ≥75% (Humidity over than 85%)
- Alarm: Audible and visual alarm; alarm events reviewable
Physical Characteristics:
- Size : 970mm×570mm×1350mm
- Screen Size :8″ TFT touch screen
- Weight :100kg
- Water Tank Capacity: ≥1L
Features :
- Air flow rate inside thermostat cover: ≤0.2m/s
- Work mode: Air/ Infant
- Bed tilt range: -10 ~ +10°
- Input voltage 220 V~
- Input current 6.6 A
- Input frequency 50/60 Hz
- Auxiliary output supply
- Output voltage 240 V
- Output frequency 50/60 Hz
- Output current 0.9A
- Oxygen Servo-Controller
- Setting range: 21~60%
- Measurement range: 10~100%
- Accuracy: ±2%
- O2 supply pressure: 0.27~0.7MPa
- O2 Concentration Calibration Available
- Weight Scale
- Measurement Range: 300g~8000g
- Resolution: 1g
- Accuracy: ±10g
- Trend View: 7days
Reviews